Zamorano Calvo, Jaime and Muñoz Marín, Víctor Manuel (2010) Calibration of SQM-L photometers for the NixNox project. [Working Paper or Technical Report] (Unpublished)
| PDF 364Kb |
Abstract
Twelve SQM-L night sky photometers have been tested and analyzed to determine their internal precision and differences in response. These photometers will be used by amateur astronomers around Spain, under the supervision of the Sociedad Española de Astronomía (SEA), to locate and characterize sites with dark skies well suited to perform astronomical observations (NixNox Project). A simple experimental setup has been built to obtain zero offsets for each photometer in order to correct all the observations.
| Item Type: | Working Paper or Technical Report |
|---|---|
| Additional Information: | This work is part of the trainee project of Víctor Manuel Muñoz as student of the International School for Advanced Instrumentation (IScAI) |
| Uncontrolled Keywords: | Astrofísica, Astrophysics |
| Subjects: | Sciences > Physics > Astrophysics Sciences > Physics > Optics Sciences > Physics > Atmospheric physics |
| Series Name: | LICA report |
| Volume: | UNSPECIFIED |
| Number: | #01 |
| ID Code: | 12262 |
| Deposited On: | 18 Feb 2011 11:13 |
| Last Modified: | 18 Feb 2011 11:13 |
Repository Staff Only: item control page