Publication:
Squareness error calibration of a CMM for quality control of ophthalmic lenses

Loading...
Thumbnail Image
Full text at PDC
Publication Date
2013-09
Advisors (or tutors)
Editors
Journal Title
Journal ISSN
Volume Title
Publisher
Springer London Ltd.
Citations
Google Scholar
Research Projects
Organizational Units
Journal Issue
Abstract
This paper proposes a calibration procedure for the squareness error of small-sized coordinate measuring machines like those used for ophthalmic lenses quality control. The proposed procedure only requires a spherical lens as a reference surface and it only needs to be measured once making this calibration method faster and cheaper than the existing ones.
Description
© The Author(s) 2013. This article is published with open access at Springerlink.com. This work was supported by the Spanish Ministry of Science and Technology under grant DPI2009-09023. This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
Keywords
Citation
1. Barakat NA, ElbestawiMA, Spence AD (2000) Kinematic and geometric error compensation of a coordinate measuring machine. Int J Mach Tools Manuf 40(6):833–850. 2. Okafor AC, Ertekin YM (2000) Derivation of machine tool error models and error compensation procedure for three axes vertical machining center using rigid body kinematics. Int J Mach Tools Manuf 40 (8): 1199–1213. 3. Ramesh R, Mannan MA, Poo AN (2000) Error compensation in machine tools—a review: part I: geometric, cutting-force induced and fixture-dependent errors. Int J Mach Tools Manuf 40(9): 1235–1256. 4. Wand BT, Bressler VE, Ruland R Squareness error mapping of a coordinate measuring machine using a two-dimensional reference object, SLAC Publication, SLAC-PUB-5836. 5. Lee ES, Burdekin M (2001) A hole-plate artifact design for the volumetric error calibration of CMM. Int J Adv Manuf Technol 17:508–515. doi:10.1007/s001700170151. 6. Liebrich T, Bringmann B, KnappW(2009) Calibration of a 3D-ball plate. Precision Engineering 33(1):1–6. 7. Kruth JP, Zhou L, Van den Bergh C, Vanherck P (2003) A method for squareness error verification on a coordinate measuring machine. Int J Adv Manuf Technol 21:874–878. doi:10.1007/s00170-002-1408-x. 8. Lagarias JC, Reeds JA, Wright MH,Wright PE (1998) Convergence properties of the nelder mead simplex method in low dimensions. SIAM J Optim 9(1):112–147.
Collections