During the last years, much interest has been dedicated to the use of amorphous silicon PIN diodes as particle and radiation detectors for medical applications. This work presents the spectrometric characterization of PECVD high deposition rate diodes fabricated at our laboratory, with thickness up to 17.5 μm. Results show that the studied devices detect the Am^(241) alpha particles and the medical X-rays generated by a mammograph model Senographe 700T from General Electric. Possible reasons of the observed energy losses are discussed in the lest. Using the SRIM2000 program, the transit of 5.5 MeV alpha particles through a diode was simulated, determining the optimum thickness for these particles to deposit their energy in the intrinsic layer of the diode.