Repeatability of Noninvasive Keratograph 5M Measurements Associated With Contact Lens Wear

Impacto

Downloads

Downloads per month over past year



García Montero, María and Rico del Viejo, Laura and Lorente Velázquez, Amalia and Martínez Alberquilla, Irene and Hernández Verdejo, José Luis and Madrid Costa, David (2019) Repeatability of Noninvasive Keratograph 5M Measurements Associated With Contact Lens Wear. Eye & Contact lens . ISSN 1542-2321

[thumbnail of Preprint]
Preview
PDF (Preprint)
667kB

Official URL: https://journals.lww.com/claojournal/Fulltext/2019/11000/Repeatability_of_Noninvasive_Keratograph_5M.6.aspx




Abstract

Objetive: To assess the intrarater repeatability of the measurements of tear meniscus height (TMH), noninvasive keratograph tear break-up time (NIKBUTs), and ocular redness measurements obtained with the Keratograph 5M (K5M) in a sample of soft silicone hydrogel contact lens (CL) wearers over 15 days.
Methods: Prospective study over two consecutive weeks. Three measurements of TMH, NIKBUTs (NIKBUT first and NIKBUT average), and ocular redness were obtained in different sessions; the first day (baseline, at 8 hr of wear, and after lens removal) and the last day of wear-15th day (at 8 hr of wear and after lens removal). The repeatability of measurements were assessed by two intraclass correlation coefficient (ICC) forms; single measurement [ICC (2,1)] and multiple measurements (k=3) [ICC (2, k)].
Results: Sixty-four eyes were analyzed. The repeatability of baseline TMH [ICC (2,1) greater than 0.90; coefficient of repeatability (CR)=0.06 mm] and after and during CL wear [ICC (2,k) greater than 0.90; CR≤0.07 mm] were excellent. The repeatability of baseline NIKBUT average [ICC (2,k)=0.89 (0.82-0.93); CR=6.07 sec] was maintained after CL removal but was poorer during CL wear. The repeatability of baseline NIKBUT first [ICC (2,k) =0.80 (0.69-0.87); CR=8.74 sec] was maintained after CL removal and during CL wear at moderate-good level.
Conclusions: Intrarater repeatability of TMH, NIKBUTs, and ocular redness performed by K5M after CL wear remains stable when three measurements are performed. However, intrarater repeatability during CL wear decreased only for NIKBUT average and was not affected by time of use (15 days).


Item Type:Article
Additional Information:

Epub ahead of print (26/04/2019)

Uncontrolled Keywords:Keratograph 5M; Repeatability; Contact lens; Tear film
Subjects:Medical sciences > Optics > Contact lenses
Medical sciences > Optics > Physiological optics
ID Code:55122
Deposited On:06 May 2019 10:37
Last Modified:05 Apr 2022 08:53

Origin of downloads

Repository Staff Only: item control page