DIAGNOSTIC PERFORMANCE OF CONE-BEAM COMPUTED TOMOGRAPHY TO DIAGNOSE IN VIVO/IN VITRO ROOT RESORPTION: A SYSTEMATIC REVIEW AND META-ANALYSIS

Impacto

Downloads

Downloads per month over past year

Baena de la Iglesia, Teresa and Yáñez Vico, Rosa María and Iglesias Linares, Alejandro (2022) DIAGNOSTIC PERFORMANCE OF CONE-BEAM COMPUTED TOMOGRAPHY TO DIAGNOSE IN VIVO/IN VITRO ROOT RESORPTION: A SYSTEMATIC REVIEW AND META-ANALYSIS. Journal of Evidence-Based Dental Practice . p. 101803. ISSN 15323382

[thumbnail of 1-s2.0-S1532338222001294-main.pdf]
Preview
PDF
Creative Commons Attribution.

2MB

Official URL: https://doi.org/10.1016/j.jebdp.2022.101803




Abstract

Background
This review analyses the diagnostic performance of cone-beam computed tomography (CBCT) for the in vivo/in vitro detection of external root resorption (ERR) and critically analyses current and past methods of measuring or classifying ERR in vivo/in vitro in terms of radiation doses and cumulative radiation risks.

Methods
A diagnostic test accuracy (DTA) protocol was used for a systematic review of diagnostic methods following PRISMA guidelines. The protocol was registered with PROSPERO (ID: CRD42019120513). A thorough and exhaustive electronic search of 6 core electronic databases was performed, applying the ISSG Search Filter Resource. The eligibility criteria were designed [problem-intervention-comparison-outcomes (PICO) statement: Population, Index test, Comparator, Outcome] and methodological quality was assessed by QUADAS-2.

Results
Seventeen papers were selected from a total of 7841 articles. Six in vivo studies were assessed as having a low risk of bias. The overall sensitivity and specificity of CBCT for diagnosis of ERR was 78.12% and 79.25%, respectively. The highest and lowest sensitivity and specificity of CBCT for diagnosis of external root resorption are 42%-98% and 49.3%-96.3%.

Discussion
Most of the selected studies reported quantitative diagnoses with single linear measurements of ERR even though multislice radiographs were available. The cumulative radiation dose (μS) to radiation-sensitive structures, such as the bone marrow, brain and thyroid, was observed to increase using the 3-dimensional (3D) radiography methods reported.

Conclusions
The highest and lowest sensitivity and specificity of CBCT for diagnosis of external root resorption are 42%-98% and 49.3%-96.3%. The minimum and maximum effective doses of dental CBCT for external root resorption diagnosis are 34 μSv and 1073 μSv.


Item Type:Article
Additional Information:

CRUE-CSIC (Acuerdos Transformativos 2022)

Uncontrolled Keywords:Accuracy; Diagnostic; External root resorption; Cone-beam computed tomography; Radiation dose
Subjects:Medical sciences > Dentistry > Dental instruments and apparatus
ID Code:76509
Deposited On:06 Feb 2023 12:07
Last Modified:06 Feb 2023 12:13

Origin of downloads

Repository Staff Only: item control page