Easyton ® transpalpebral versus Perkins applanation tonometry in different clinical populations

Loading...
Thumbnail Image
Full text at PDC
Publication Date
2023-01-11
Authors
Montolío Marzo, Elena
Morales Fernández, Laura
Sáenz Francés, Federico
García Sáenz, Sofía
Piñero, David Pablo
Martínez de la Casa, Jose Maria
Advisors (or tutors)
Editors
Journal Title
Journal ISSN
Volume Title
Publisher
Citations
Google Scholar
Research Projects
Organizational Units
Journal Issue
Abstract
Objective: To compare intraocular pressure (IOP) measurements obtained using the new transpalpebral Easyton® tonometer and Perkins applanation tonometer (PAT) in three different clinical populations. Methods: The participants of this prospective study were 84 subjects divided into the groups: 22 healthy children (G1), 42 healthy adults (G2), and 20 adult patients with primary open angle glaucoma (G3). The data recorded in 84 eyes of these subjects were age, sex, eye, central corneal thickness (CCT) and axial length (AL). In all eyes, IOP was determined in the same examination room by the same experienced examiner using Easyton® and PAT in random order. Results: Mean differences in IOP readings between Easyton® and PAT were 0.45 ± 1.97 (p = 0.295), -0.15 ± 2.13 (p = 0.654), -1.65 ± 3.22 (p = 0.033), and − 0.018 ± 2.500 mmHg (p = 0.500) in the groups G1, G2, G3 and whole sample (G4), respectively. Correlations between Easyton® and PAT IOP values were 0.668 (p = 0.001) for G1, 0.463 (p = 0.002) for G2, 0.680 (p < 0.001) for G3 and 0.605 (p < 0.001) for G4. Moderate to good agreement between the two tonometers was found in all groups according to intraclass correlation coefficients, which were 0.794 (p < 0.001) for G1, 0.632 (p < 0.001) for G2, 0.809 (p < 0.001) for G3, and 0.740 (p < 0.001) for G4. The lower and upper limits of agreement between the devices were − 5.1 and 4.7 mmHg, respectively. No correlation was noted between CCT or AL and the Easyton® IOP measurements. Conclusion: IOP measurements obtained with Easyton® and PAT show an acceptable level of agreement both in healthy individuals and in patients with glaucoma.
Description
UCM subjects
Keywords
Citation
Collections