Three-year follow-up of posterior chamber phakic intraocular lens with a central port design after deep anterior lamellar keratoplasty

Impacto

Downloads

Downloads per month over past year

Alfonso Bartolozzi, Belén and Lisa Fernández, Carlos and Fernández Vega-Cueto, Luis and Baamonde Arbaiza, Begoña and Madrid Costa, David and Alfonso Sánchez, José F. (2022) Three-year follow-up of posterior chamber phakic intraocular lens with a central port design after deep anterior lamellar keratoplasty. Eye and Vision, 9 (1). art.34-9 p.. ISSN 2326-0254

[thumbnail of s40662-022-00306-1.pdf]
Preview
PDF
Creative Commons Attribution.

901kB

Official URL: https://doi.org/10.1186/s40662-022-00306-1




Abstract

Background: To evaluate clinical outcomes of the Visian implantable collamer lens (ICL) with a central port to correct myopia and astigmatism after deep anterior lamellar keratoplasty (DALK) for keratoconus throughout 3 years of follow-up.
Methods: This study included 20 eyes of 20 patients that underwent V4c ICL (13 eyes with a spherical ICL and 7 eyes with a toric ICL) implantation after DALK. Uncorrected (UDVA) and corrected (CDVA) distance visual acuities, refraction, intraocular pressure (IOP), endothelial cell density (ECD), and vault were analyzed.
Results: The mean UDVA improved from the preoperative 1.18 ± 0.33 logMAR to 0.25 ± 0.14 logMAR at 6 months after surgery (P < 0.0001) and remained unchanged throughout the whole follow-up (P = 0.4). All eyes gained lines of CDVA compared to preoperative values. At the last follow-up visit, all eyes achieved CDVA of 0.2 logMAR or better and 13 eyes (65%) 0.1 logMAR or better. At 6 months post-surgery, all eyes (100%) had a spherical equivalent within ± 1.50 D, and 19 (95%) within ± 1.00 D. The mean manifest spherical equivalent was stable over the postoperative follow-up (P = 0.25). No significant increase in IOP occurred in any case throughout the 3 years of follow-up. The loss in ECD from the preoperative baseline at the last follow-up visit was 2.27%.
Conclusions: The clinical outcomes suggest that the V4c ICL implantation for correction of myopia and regular astigmatism in post-DALK eyes was satisfactory in terms of effectiveness, safety, and stability during 3 years of follow-up.


Item Type:Article
Uncontrolled Keywords:DALK; Deep anterior lamellar keratoplasty; ICL; Implantable collamer lens.
Subjects:Medical sciences > Medicine > Surgery
Medical sciences > Medicine > Ophtalmology
Medical sciences > Optics > Optometry
ID Code:77685
Deposited On:19 May 2023 18:02
Last Modified:19 May 2023 18:02

Origin of downloads

Repository Staff Only: item control page